Instructions for Use
41
8
Maintenance and Servicing
Although the Co-mix unit does not require any scheduled servicing, the
operator should regularly clean the unit and inspect it for any detects, as
described in section 8.2 below.
DANGER
WARNING
Please observe and comply with all of the
Unit Maintenance
and Serviceability
precautions listed in section 2.
Never removed the unit casework. There are no user or
operator serviceable parts inside the unit.
Always switch off and unplug the unit before performing any
cleaning or disinfecting tasks.
For technical and service related enquiries, please contact your distributor or
ITL at the address given on page 2 of this manual.
8.1
Replacing the Unit Fuse
The unit fuse should only be replaced by a suitably qualified engineer.
DANGER
The unit fuse will only blow as a result of an internal unit fault.
This fuse should only be changed after the unit has been
thoroughly inspected by a qualified engineer, and must be
replaced with the exact type specified in section 11.
Thoroughly inspect the unit for any signs of damage, loose components or
liquid spillage or ingress. If in doubt, please contact ITL on the number given
on page 2 of this manual.
DANGER
The fuse holder is removed by disconnecting the mains power
cord and then using a flat bladed screwdriver to carefully pry
open the fuse access cover and ease out the fuse holder.
After replacing the fuse with an identically rated fuse (see
section 11), push the fuse holder firmly back into the inlet
module and close the access cover.
The unit must be electrically safety tested for excess leakage
current before being repowered from the mains supply.