Instructions for Use
21
6.3.2
Pre-defined Mixing Programs
The unit has eight pre-defined mixing programs which support the commonly
used plates and micro test tubes listed in Table 7.
These programs are listed in Table 8, along with details of the pre-set mixing
speed and time parameters, the recommended well or tube fill level and the
required adaptor block type.
Table 8: Pre-defined Mixing Programs
Program
ID
Plate or Tube
Type
Speed
(RPM)
Time
(seconds)
Fill
Level
Adaptor
Block
MTP-96
96 well MTP
1000
30
5 to 60%
x
DWP-96
96 well DWP
1650
30
5 to 50%
x
PCR-96
96 well PCR plate
0.2ml tubes or strips
1650
30
5 to 50%
1
MTP-384
384 well MTP
2000
15
10 to 60%
x
DWP-384
384 well DWP
2000
15
10 to 60%
x
PCR-384
384 well PCR plate
2600
15
10 to 50%
x
Micro 0.5
0.5ml micro tubes
1650
30
5 to 50%
2
Micro 1.5/2.0 1.5ml micro tubes
2.0ml micro tubes
1400
60
5 to 80%
3
Other mixing speeds and times can be selected using the free-
style MIX program (see section 6.3.1) or by creating a new user-
defined program (see section 6.3.5 for details).
The procedure for mixing using a pre-defined program is similar to that
described in section 6.3.1 for the free-style MIX program.
First select the appropriate pre-defined program from the selection grid, and
then press the
Accept
button to load it.
→
→
→
Содержание Co-mix
Страница 1: ...Instructions for Use...