Instructions for Use
39
To change a setting, press the associated option button, then select the new
value, and press the
Accept
button to save the new setting.
Cancel
Accept
7.4.1
Beeper and Button Click Volume
The beeper is sounded at the end of each mixing or vortexing operation. This
volume can be adjusted to suit the ambient sound conditions, or muted.
The touchscreen button click volume can also be independently set or
permanently
Disabled
.
7.4.2
LCD Backlight Brightness Level
The LCD backlight brightness can be adjusted to suit different lighting
conditions.
7.4.3
Cooling Fan Operating Mode
The unit’s internal cooling fan draws air in from around the plate holder and
gently blows warm air out of the ventilation slots in the base of the unit.
It is recommended that the mode is set to
Automatic
, which operates the fan
only when required. However, if the unit is being used in a temperature-
controlled chamber, it may be desirable to ensure consistence airflow by
setting the cooling fan to run
Continuous
or turning it
Off
completely.
7.4.4
Program Set-up Protection Mode
By default, the user is able to access all features and functions of the unit.
However, once the unit has been installed and fully set-up, it may be
desirable to disable these programming functions to protect against
accidental modification.