Criticare Systems, Inc.
VitalCare
TM
506N3 Series
Service Manual
Page 7-7
Section 7 —Disassembly
4. Remove the four (4) screws (pn 40995B005) from the
isolation PCB.
Figure 7-7: Remove Isolation Board Screws
5. Carefully lift up the isolation PCB (pn 91386A001).
6. Remove the Temp cable (pn 90931A001) from the connector
beneath the PCB.
NOTE
: Do not tuck temp cable into opening.
7. Flip the PCB over and disconnect the black and red wire from
J3
of the isolation board. (This cable is soldered to the FasTemp
PCB.)
Figure 7-8: Isolation Board Connectors
8. Disconnect the RS232 cable (pn 90933A001) from
J4
on the
isolation PCB.
9. Disconnect the flex cable (pn 83266B002) from
J2
by unlocking
the tab and pulling it out.
NOTE
: Make sure that the contacts will face up when reinserting
the cable.
10.Replace the isolation PCB.
11.Reassemble in reverse order.
Follow “Completion of Service” procedure at the end of this section.
Remove Screws (torqued @ 5 in lbs)
Temp cable
underneath PCB
J1
J3
J4
J2
Содержание 506DN3
Страница 12: ......
Страница 56: ......
Страница 62: ......
Страница 90: ......
Страница 156: ......
Страница 184: ......
Страница 225: ......
Страница 229: ......
Страница 230: ......
Страница 231: ......
Страница 232: ......
Страница 233: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 241: ......
Страница 242: ......
Страница 243: ......
Страница 244: ......
Страница 245: ......
Страница 247: ......
Страница 248: ......
Страница 249: ......
Страница 250: ......
Страница 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Страница 253: ......
Страница 254: ......