Section 5 —Preventative Maintenance
Page 5-24
VitalCare
TM
506N3 Series
Service Manual
Criticare Systems, Inc.
13.Remove the probe. The LED should flash by the legs on the
human icon.
14.Place the probe into a probe cover and place the probe under
an armpit to verify a temperature reading appears. Reinsert the
probe into the chamber/well.
Optional: Other temperatures and functions may be verified with the
Alaris TurboTemp Probe Simulator. (TurboTemp models only.)
Simulator Setting
Thermometer Reading
80.2
80.2
±0.2 F (26.7 ±0.1 C)
98.0
98.0
±0.2 F (36.7 ±0.1 C)
98.6
98.6
±0.2 F (37.0 ±0.1 C)
102.0
102.0
±0.2 F (38.9 ±0.1 C)
107.8
107.8
±0.2 F (42.1 ±0.1 C)
B.P.
106.0
±0.2 F (41.1 ±0.1 C)
Verify
TEMP: BAD PROBE
message displays when the Broken Probe
button is pressed on IVAC simulator when the dial is set for B.P.