Criticare Systems, Inc.
VitalCare
TM
506N3 Series
Service Manual
Page 5-19
Section 5 —Preventative Maintenance
Monitoring Module
Verification
NIBP Verification
This applies to all 506N3 Series monitors. The NIBP verification
requires Dynatech Nevada NIBP Analyzer. Connect the 506N3 Series
monitor to a Dynatech Nevada NIBP Analyzer set for the following
operation.
NIBP Analyzer Settings
Adult 120/80 (90); Heart Rate 120 bpm
Pressure Adjustments: Gain 100%; Shift 4
Use a tee connection with an adult dummy cuff. Connect to the 0-300
mmHg port of the NIBP analyzer. A neonatal dummy cuff is also
required for complete testing.
It is recommended that the actual cuffs (to be used with the monitor)
are setup as dummy cuffs for this verification. Wrap the cuff snug
around a sturdy cylinder.
1. Connect the monitor to an AC power source and turn on the
monitor. Set the monitor to the
ADULT
Mode (Adult Patient) by
pressing the PATIENT key until the ADULT icon illuminates
.
2. Press the NIBP START/STAT/STOP key and allow the monitor to
take at least four (4) readings. The systolic, diastolic and mean
readings should not vary by more than ±4% or ±4mmHg
(whichever is greater) from the calculated average. Each
reading shall not vary more than 8 mmHg from the simulator
setting.
Leave the monitor connected to the NIBP analyzer. Change the NIBP
analyzer setting to the neonate configuration with a simulator setting
of 80/50 (62) mmHg; Heart Rate 80 bpm. A neonatal dummy cuff
must be used during this test.
1. Set the monitor to
NEONATE
Mode by pressing the PATIENT
key until the NEONATE icon illuminates.
2. Press the NIBP START/STAT/STOP key and allow the monitor to
take at least four (4) readings. The systolic, diastolic and mean
readings should not vary by more than ±4% or ±4mmHg
(whichever is greater) from the calculated average. Each
reading shall not vary more than 8 mmHg from the simulator
setting.
3. If the monitor continues to fail verification contact the Criticare
Technical Support Department.
Patient
Patient
Содержание 506DN3
Страница 12: ......
Страница 56: ......
Страница 62: ......
Страница 90: ......
Страница 156: ......
Страница 184: ......
Страница 225: ......
Страница 229: ......
Страница 230: ......
Страница 231: ......
Страница 232: ......
Страница 233: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 241: ......
Страница 242: ......
Страница 243: ......
Страница 244: ......
Страница 245: ......
Страница 247: ......
Страница 248: ......
Страница 249: ......
Страница 250: ......
Страница 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Страница 253: ......
Страница 254: ......