Section 7 —Disassembly
Page 7-32
VitalCare
TM
506N3 Series
Service Manual
Criticare Systems, Inc.
4. Remove the four(4) remaining screws located on the top,
bottom left and middle. (Or three (3) screws on the top and
middle if the monitor had a ground wire which you removed in
the previous step.)
Figure 7-40: Remove Screws
5. Unlock the membrane tail at
J11
and remove the tail from the
connector.
6. Remove the assembly from the bezel.
NOTE
: For monitors with a Bezel Support (pn 45190B001),
retain the Bezel Support.
7. Flip the assembly over. Determine which assembly needs to be
replaced.
• If the LCD board needs to be replaced, remove the four (4)
screws (pn 41258B003) using a #0 Phillips screwdriver.
• If the LED board needs to be replaced, remove the six (6)
screws (pn 40995B005) using a #1 Phillips screwdriver.
• Carefully attach either the new LCD board (pn 90029A001) or
LED display board (91388A001).
Remove
Screws
J11
Содержание 506DN3
Страница 12: ......
Страница 56: ......
Страница 62: ......
Страница 90: ......
Страница 156: ......
Страница 184: ......
Страница 225: ......
Страница 229: ......
Страница 230: ......
Страница 231: ......
Страница 232: ......
Страница 233: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 241: ......
Страница 242: ......
Страница 243: ......
Страница 244: ......
Страница 245: ......
Страница 247: ......
Страница 248: ......
Страница 249: ......
Страница 250: ......
Страница 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Страница 253: ......
Страница 254: ......