Section 7 —Disassembly
Page 7-4
VitalCare
TM
506N3 Series
Service Manual
Criticare Systems, Inc.
6. Lift the printer assembly up from the monitor.
Figure 7-3: Lift Printer Assembly
7. Label and remove the printer ribbon cable (pn 90928A001) from
the connection on the printer assembly.
8. Attach the printer ribbon cable to the new printer assembly.
9. Set the new printer assembly on the top of the monitor. Press
the assembly until it snaps into place.
10.Fasten the printer assembly to the monitor with the two (2)
painted screws removed earlier.
NOTE:
While securing the two (2) screws, press down on the
rear of the printer housing to keep the printer flush with the
monitor’s rear housing.
11.Reconnect the battery as described in “Battery Removal/
Replacement” in Section 5.
12.Perform a functional test of the printer as described in
“Functional and Safety Testing” in Section 5.
Remove Ribbon Cable