Section 5 —Preventative Maintenance
Page 5-4
VitalCare
TM
506N3 Series
Service Manual
Criticare Systems, Inc.
REPLACE THE BATTERY
1. Attach the battery cables to the new battery (pn 80518B001).
IMPORTANT
: Connect the red battery cable to the positive
battery terminal and the black cable to the negative battery
terminal.
Figure 5-3: Connecting the Battery
2. Slide battery into the battery bay.
3. Reattach the battery cover with the four (4) screws removed
earlier.
Fuse Removal/
Replacement
There are two AC power fuses located at the rear of the monitor
directly below the AC power entry socket.
• Use 0.5A 250V time delay fuses (T 500MA L 250V).
1. With a flat blade screwdriver, turn the fuse cover(s) out.
2. Gently pull the fuse cover(s) with fuse(s) out of the fuse
assembly.
Figure 5-4: 506N3 Series Fuses
3. Gently pull the fuse(s) out of the fuse cover(s).
4. Reassemble in reverse order.
RED
BLK
Fuse Covers
Fuses
Fuse Wells
Содержание 506DN3
Страница 12: ......
Страница 56: ......
Страница 62: ......
Страница 90: ......
Страница 156: ......
Страница 184: ......
Страница 225: ......
Страница 229: ......
Страница 230: ......
Страница 231: ......
Страница 232: ......
Страница 233: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 241: ......
Страница 242: ......
Страница 243: ......
Страница 244: ......
Страница 245: ......
Страница 247: ......
Страница 248: ......
Страница 249: ......
Страница 250: ......
Страница 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Страница 253: ......
Страница 254: ......