Criticare Systems, Inc.
VitalCare
TM
506N3 Series
Service Manual
Page 2-3
Section 2 —Service Menus
BOARD
SETUPS
SpO
2
NONE
Selects which type of SpO
2
module is installed in
monitor. Choices are
NONE
,
DOX
(Criticare
Digital Oximetry), and
NCOR
(Nellcor)
NIBP
1020
Selects which type of NIBP module is installed in
monitor. Choices are
NONE
and
1020
.
TEMP
NONE
Selects which type of Temperature module is
installed in monitor. Choices are
NONE, FSTMP
(FasTemp) and
Turbo
(TurboTemp).
DEFAULT
SETUPS
CONFIG USER
USER
Selects which type of configuration defaults are
restored when the MENU key is held on power up.
Choices are
USER
,
HOSP.
(hospital),
ALT C.
(alternate care).
STORE USER
NO
• Allows the storing of current configuration
settings as USER DEFAULTS. Choices are
NO
and
YES
.
• Pressing the MENU key with
YES
displayed
causes this line in the LCD to be replaced by
CONFIRM
with the choices of
NO
and
YES
.
• Pressing the MENU key with
YES
displayed
shall store to memory.
DONE
replaces
YES
in
the selection upon completion.
AUDIO OFF
YES
Selects the nature of Alarm Volume setting of
OFF
in the
MAIN MENU
,
ALARMS
configuration.
•
YES
indicates true silencing of the audio alarm.
•
NO
causes the audio to not annunciate alarms
but sounds a double beep every two minutes for
verification that the audio circuit still functions.
LINE FREQ
60
The monitor has a
60
Hz frequency setting for
domestic U.S. use and a
50
Hz setting for
international use. The frequency must be set
correctly to the local AC (Mains) power frequency
for the monitor to function correctly. Contact your
distributor for more information about which
setting to use.
TEST MENU
LED TEST
OFF
Tests all segments of the numeric LED display
and the single LEDs. Choices are
OFF
and
ON
.
Press the MENU key with
ON
displayed to begin
the test. This test cannot be canceled once
started. The message
DONE
replaces
ON
in the
selection when the test is completed.
PRINT TEST
OFF
Tests the internal printer by printing all printable
characters on the printer paper. Choices are
OFF
and
ON
. Press the MENU key with
ON
displayed
to begin the test. The message
DONE
replaces
ON
in the selection when the test is completed.
NIBP SEAL
OFF
Tests NIBP pressure seals and transducer.
Choices are
OFF
and
ON
. Press the MENU key
with
ON
displayed to begin the test sequence with
a new prompt message.
Primary Menu
Secondary Menu
Default
Description
Содержание 506DN3
Страница 12: ......
Страница 56: ......
Страница 62: ......
Страница 90: ......
Страница 156: ......
Страница 184: ......
Страница 225: ......
Страница 229: ......
Страница 230: ......
Страница 231: ......
Страница 232: ......
Страница 233: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......
Страница 241: ......
Страница 242: ......
Страница 243: ......
Страница 244: ......
Страница 245: ......
Страница 247: ......
Страница 248: ......
Страница 249: ......
Страница 250: ......
Страница 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Страница 253: ......
Страница 254: ......