Command
Function
[SPM_ARCT]
CT mode
[SPM_CSON]
CT sensor power on
[SPM_FISS_0001]
Normal Mode : 4x4 binning & High gain
[SPM_FIPM_0001]
Sensor size setting
[SPM_FISM_0001]
Sensor clock (1:External, 0: Internal)
[SPM_FREQ_0100]
Sensor frequency 100 Hz
[SPM_COLM_0004]
AL filter position (2 Point)
If Al Filter is not completely covered, adjust with [SPM_CEST_XXXX]
[SPM_COLM_0010]
1T Cu + AL filter location (4 Point)
If 1T Cu + Al Filter is not all covered, send [SPM_CBST_XXXX] to adjust.
[SPM_AAMS]
Uoffset driven
[SPM_ABMS]
FOV 8x8 driven