Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
81 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
4.3.2. Rotator Y, Z-axis setting
1. Put the chin block and the normal bite on the unit as shown below.
2. Run
VAKCAP.exe
file.
3. Click the
Capture/Recon
tab and send the
[SPM_ARCT]
command to initiate the system into
CT mode.
4. When the equipment stops operating, send the following commands in the table below to supply
power to the CBCT Sensor.
Command
Function
[SPM_CSON]
CBCT sensor Power On
[SPM_FISS_0001]
Sensor Sensitivity & Binning mode setting
//0000: High Gain, 2x2 Binning
//0001: High Gain, 4x4 Binning
//0002: Low Gain, 2x2 Binning (default)
//0003: Low Gain, 4x4 Binning
[SPM_FIPM_0001]
Sensor Partial mode setting
// Sensor size: 0000~0032
[SPM_FISM_0001]
Sensor Internal/External mode setting
//0000: Internal
//0001: External
[SPM_COLM_0004]
Collimator Position set
5. Send the [SPM_ABMS] command to acquire dark.
6. Click the
Soft Calib
tab and click the
Dark
button to acquire Dark Calibration Data.