3.5.
PANO sensor: Bright calibration
•
Stay outside of the X-ray shielding room during Bright Calibration data
achievement.
1. Run the C:\VCaptureSW\Acquisition\Pano\WidePANO\VACAP.exe file.
2. Click the
Settings
tab and ensure that
[BrightCalibration]
kVp and mA are set
3. When acquiring 5 Point cal, the initial value is as follows, and it is set according to the standard
by adjusting kVp.
Acquisition conditions
5 point
exposure
condition
kVp0
570
mA0
550
kVp1
850
mA1
870
1. If the standard is not satisfied, set them according to the
standard by adjusting kVp0 and kVp1.
2. Click the "Bright" button to obtain 5 points of Bright Cal Data and
check if the following standard is met.
4. Send the
[SPM_PSON]
command to supply power to the PANO Sensor.