3.5.2. Bright calibration: 4 points
1. Run the C:\VCaptureSW\Acquisition\CBCT\WidePANO\VACAP.exe file.
2. Click the
Settings
tab and make sure that the parameters under [BrightCalibration] are as
shown below and click the
Save
button if something is changed.
3. Click the
Soft Calib
tab.
4. When acquiring 4 point cal, the initial value is as follows and it is set according to the
standard by adjusting kVp.
Acquisition conditions
Half Scan
4 point
2 point
exposure
condition
kVp0
580
710
mA0
530
700
kVp1
920
830
mA1
950
900
5. Send [SPM_ARCT], [SPM_CSON], [SPM_FIPM_0001], [SPM_FISS_0001],
[SPM_FISM_0001], [SPM_COLM_0010] commands to the command window to set the CT
sensor and collimator.