4.
Error Message List
•
Tube Error Message Sent From MCU To PC
Command
Description
[SMP_ERRR_0001]
The tube is not ready
[SMP_ERRR_0002]
Tube Mono Block and Inverter Board Cable are not connected
[SMP_ERRR_0003]
Current is more than maximum permissible current while Inverter
Board is working (X-Ray exposure)
[SMP_ERRR_0004]
This error occurs when there is a problem with the tube voltage
Ref. This error occurs when there is a difference of + -10 kV or
more in the Ref. normal value.
[SMP_ERRR_0005]
This error occurs when there is a problem in the tube current Ref.
and there is a difference of +-0.5mA or more in the Ref. normal
value.
[SMP_ERRR_0006]
An error occurs when there is a problem with the tube voltage
feedback. This error occurs when there is a difference of + -20kV
or more in the normal value of Feed Back.
[SMP_ERRR_0007]
An error occurs when there is a problem with the tube's current
feedback. This error occurs when there is a difference of + -1 mA
or more in the normal value of Feed Back.
[SMP_ERRR_0008]
The temperature of Mono Block is over 55
℃
[SMP_ERRR_0009]
The current of High Voltage(EP, IP) is more than the permissible
current during X-Ray exposure.
[SMP_ERRR_0010]
Although X-Ray Switch is turned on, X-Ray On Command is not
send in the System
[SMP_ERRR_0011]
Although X-Ray Switch is turned off, X-Ray Off Command is not
sending in the System
[SMP_ERRR_0012]
During X-Ray exposure, kVp Feedback is less than -20 kVp off
from the setting value.
Error “3” is displayed on Inverter Board.
[SMP_ERRR_0013]
During X-Ray exposure, kVp Feedback is more than +20 kVp off
from the setting value.
Error “4” is displayed on Inverter Board.
[SMP_ERRR_0014]
During X-Ray exposure, mA Feedback is less than 50% off from