Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
52 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
2.
Equipment default values
2.1. Check the equipment horizontal level
1. Check the equipment level at the same position as the picture above (standard: ± 0.1° or less).
2. At the position where the sensor is looking at the column, visually check whether the vertical and
the part marked by the red dotted line of the rotator are parallel.
3. If it is not parallel, send the [SPM_ ACHF_XXXX] command to set the CT Mode Half Value to
make it parallel, and input the same value to PANO Mode Half Value [SPM_HFST_XXXX]
command.
•
As the value increases, the sensor moves to the right.