Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
164 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
10. Complete the CEPH Calibration, and choose the
①
CEPH Processing button, as shown in [8].
11. When the Mode Confirm button is chosen, Level 1~3 images are created.
* Level 1~3: Level 3 has the highest contrast due to the difference in contrast.
12. Choose the O mark on the far right of
③
, which is an image of Level 3 among the created
images.
13. When the Simulation button is chosen, an image of O marked in red, as shown in [9], is created.
14. Choose O marked in red and
⑥
click the
Apply
button.
15. When a pop-up window is created, as shown in [10],
⑦
click the
OK
button.
16. As shown in [11], check that the time of the modified date of Acquisition/WideCEPH /Norm or
Fast/UseParam/ZeusCEPH is the same as the time when the work is completed.