COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
5. Click the
Soft Calib
tab and click the
Dark
button to acquire Dark Calibration Data.
6. Enter the (HFST value worked in step 2.3) + 3600.
ex. If it is [SPM_HFST_200], input [SPM_HFST_3800].
7. Click the
Capture_CTB
button to acquire an image.
8. Click the
V Proj
button to check the image.
9. Enter
92
on the X-axis and check if the three pins are aligned in a line as shown in the image
below.