Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
109 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
7.
Low Contrast Resolution evaluation
7.1. Low Contrast Resolution standards
Parameter
LSL (Lower Specification Limit)
USL (Upper Specification Limit)
Low Contrast Resolution
-
9.0 mm
7.2. Phantom Specification
S&C Phantom
7.3. Test Method
1. The evaluation is conducted in the same manner as in the evaluation method in Section 6.3.
7.4. Test Result
Parameter
LSL
Measurement value
USL
Pass/Fail
Low Contrast Resolution
-
8 mm