More Information
‣
Configuring Analog Outputs with Insight (page 40)
2.2.4 Installation Type
MGP261 can be installed either directly into the process (for example, through a flange or ball
valve in the actual process pipeline), or using a flow-through adapter (for example, in a
sampling line installation or when performing a field calibration).
MGP261 uses different internal calculation models depending on the installation type and gas
flow rate. For this reason, the type of the installation must be set correctly in MGP261 settings.
You can view and change the installation type configuration with Vaisala Insight PC software:
for instructions, see
Installation Type Configuration in Insight (page 33)
.
Always ensure that the installation type configuration is set
correctly in Insight when changing from one installation type to another, so that
the calculation model in use matches the installation environment and
measurement accuracy is not affected.
CAUTION!
2.2.5 Process Flow Range and Installation Type
A process flow range of 2 ... 20 m/s is suitable for the in situ installation calculation model.
When operating the probe in a low flow environment (0 … 2 m/s) or using the flow-through
adapter, use the flow-through calculation model. To set the correct calculation model for the
flow range, choose the installation type in the Vaisala Insight PC software.
Process Flow Range
Installation Type Selection in Insight PC Software
2 ... 20 m/s
Directly in process
0 … 2 m/s or with flow-through adapter
Flow-through adapter
More Information
‣
Installation Option: Ball Valve Installation (page 22)
‣
Installation Option: Flow-Through Adapter Installation (page 23)
‣
Installation Type Configuration in Insight (page 33)
‣
Connecting to Insight Software (page 35)
‣
2.2.6 Connectivity to Vaisala Insight Software
The probe can be connected to Vaisala Insight software using a Vaisala USB cable (order code:
257295). With the Insight software, you can:
• Calibrate and adjust the measurement.
• See device information and status.
Chapter 2 – Product Overview
11
Содержание MGP261
Страница 8: ...MGP261 User Guide M212273EN A 6...
Страница 50: ...4 Store the setting with Save MGP261 User Guide M212273EN A 48...
Страница 86: ...MGP261 User Guide M212273EN A 84...
Страница 88: ...MGP261 User Guide M212273EN A 86...
Страница 89: ......
Страница 90: ...www vaisala com...