7.1.3 Compensation Mode During Calibration
When you start to calibrate the probe with Insight, the probe enters calibration mode, and the
compensation mode is automatically switched to
Setpoint
. In this compensation mode, you
must enter the conditions of your calibration environment as temporary setpoint values in the
Compensation setpoints
tab of the calibration menu.
When you exit calibration mode, the values you have entered in the
Compensation setpoints
tab remain in use as the current setpoint values, but the compensation mode switches back to
the selection that was in place before starting calibration (either
Off
,
Setpoint
,
Measured
, or
External
).
More Information
‣
Configuring Environmental Compensations with Insight (page 44)
7.2 Configuring Environmental Compensations
with Modbus
The Modbus implementation of MGP261 includes configuration registers for pressure and
temperature compensation setpoints. You can also configure the temperature and pressure
compensation mode (
External
,
Setpoint
,
Measured
(temperature only), or
Off
).
For a description of MGP261 Modbus registers, see
More Information
‣
‣
7.3 Configuring Environmental Compensations
with Insight
Using Insight, you can configure the following temperature and pressure compensation
settings:
• Temperature or pressure compensation mode (select the source where the compensation
value is received from)
• The setpoint values for temperature or pressure compensation
• The scaling of the analog input channel (
Analog input 1
) that is used to receive
temperature or pressure values from an external sensor.
describes the different environmental compensation selections in Insight.
MGP261 User Guide
M212273EN-A
44
Содержание MGP261
Страница 8: ...MGP261 User Guide M212273EN A 6...
Страница 50: ...4 Store the setting with Save MGP261 User Guide M212273EN A 48...
Страница 86: ...MGP261 User Guide M212273EN A 84...
Страница 88: ...MGP261 User Guide M212273EN A 86...
Страница 89: ......
Страница 90: ...www vaisala com...