3.1.4 Installation Option: Ball Valve Installation
The following figure
shows an example MGP261 ball valve installation. The figure highlights
the correct
wiring routes and shows the recommended installation depth and orientation.
Always use a separate cable for the optional (Ex ia) external
pressure or temperature sensor input wiring (1), and make sure the wiring
remains separated from the terminals and wiring (2) on the other side of the
metal barrier (3) on the component board.
CAUTION!
1
2
4
3
5
Figure 8 MGP261 Installation Example with Wiring Routes and Recommended Orientation and
Depth
1
Optional external pressure or temperature sensor input wiring (Ex ia): use the M16x1.5
lead-through and route the cable directly to the terminal.
2
Standard analog output, power supply input, and RS-485 communication wiring: use the
M20x1.5 lead-throughs and route the cables to the terminals through the openings below
the component board.
3
Metal barrier separating the intrinsically safe external sensor input terminals (optional, for
environmental compensation input) from the analog output, power supply input, and
RS-485 terminals on the component board.
4
1.5" male NPT thread on MGP261:
never install to any other thread type than 1.5" female
NPT
.
5
For best results, install MGP261 horizontally and position the tip of the filter within 1/3 of
the pipe's diameter from the pipe centerline. In smaller pipes, the installation depth can be
adjusted by using an adapter (for example, a ball valve).
More Information
‣
Recommended Installation Position on Pipeline (page 24)
MGP261 User Guide
M212273EN-A
22