Property
Methane CH
4
Carbon Dioxide CO
2
Water Vapor H
2
O
Response time with
flow-through adapter
90 s at
≥
0.5 l/min
(recommended: 0.5 … 1 l/min)
1)
Excluding cross-interferences to other gases.
2) Time to first reading
3) Time to specified accuracy
4) At +20 °C (+68 °F) ambient temperature
5) With standard PTFE filter
Table 8 Operating Environment
Property
Specification
Operating temperature range
-40 ... +60 °C (-40 ... +140 °F)
Operating humidity range
0 ... 100 %RH
Storage temperature range
-40 ... +60 °C (-40 ... +140 °F)
Storage humidity range
0 ... 90 %RH
Process pressure range
-500 ... +500 mbar(g)
Process temperature range
+0 ... +60 °C (+32 ... +140 °F)
Process flow range
0 … 20 m/s
1)
A process flow range of 2 … 20 m/s is suitable for in situ installations (probe installed
directly into process). For process flow ranges between 0 … 2 m/s, a flow-through
installation is recommended.
Table 9 Compliance
Property
Specification
Electromagnetic compatibility (EMC)
EN61326-1(2014), Industrial environment
Ex classification
Ex II 1/2 (1) G Ex eb mb [ia] IIB T3 Ga/Gb -40 °C
≤
Tamb
≤
+60 °C
IP rating
IP66
Table 10 Inputs and Outputs
Property
Specification
Operating voltage
18 ... 30 VDC
Power consumption
Typical: 3 W
Maximum: 6 W
Chapter 10 – Technical Data
63
Содержание MGP261
Страница 8: ...MGP261 User Guide M212273EN A 6...
Страница 50: ...4 Store the setting with Save MGP261 User Guide M212273EN A 48...
Страница 86: ...MGP261 User Guide M212273EN A 84...
Страница 88: ...MGP261 User Guide M212273EN A 86...
Страница 89: ......
Страница 90: ...www vaisala com...