3.6 Attaching Flow-Through Adapter to MGP261
The MGP261 flow-through adapter 258877 is delivered assembled and is intended to be used
with the probe installed in an upright position. For an overview of the flow-through adapter,
Installation Option: Flow-Through Adapter Installation (page 23)
.
To attach the adapter to MGP261:
1. Loosen the 2 screws on the clamp.
2. Loosen the probe gland.
Only rotate the tightening nut of the probe gland. Never turn
from the locknut of the probe gland (below the tightening nut). The
threads of the locknut are glued onto the adapter with a sealant, and
rotating the locknut will compromise the leak tightness of the adapter.
CAUTION!
3. Insert the probe head into the adapter through the clamp and probe gland. Push the
probe into the adapter until the shoulder of the probe head meets the slot inside the
adapter and stops. The probe filter should cover both the inlet and outlet ports from the
inside. Check the installation depth measurement as shown in
4. Tighten the probe gland.
Chapter 3 – Installation
31