1
2
4
3
Figure 21
Measurement
Menu View in Insight PC Software
1
Compensation power-up defaults
: configure the setpoint compensation values that are
taken into use at device reset. Used when the temperature or pressure compensation
mode is set as
Setpoint
.
2
Compensation setpoints
: configure the temporary setpoint compensation values that are
in use while the device is powered on, but revert back to power-up defaults when the
device is reset. Used when the temperature or pressure compensation mode is set as
Setpoint
.
3
Measurement
(shown in figure): configure the pressure and temperature compensation
modes (available options:
Setpoint
,
External
,
Measured
(temperature only), and
Off
.
4
Analog input 1
: configure the input from the external pressure or temperature sensor
(used when the pressure or temperature compensation mode is set as
External
).
1. Connect to Insight (see
Connecting to Insight Software (page 35)
2. Select
> Configure Device
.
3. Select one of the compensation modes in the
Measurement
menu.
4. See the separate instructions for configuring the compensations in each compensation
mode:
•
Configuring Setpoint Values for Compensations with Insight (page 46)
•
Using Probe Measurement as Temperature Compensation (page 46)
•
Using Compensation Received from External Temperature or Pressure Sensor
•
Configuring Input from External Analog Pressure or Temperature Sensor (page 47)
More Information
‣
Compensation Mode During Calibration (page 44)
Chapter 7 – Environmental Compensation
45