3.2 Preparing Probe for Installation
The following tools are required when installing:
• Adjustable wrench (or a similar suitable tool) for turning the tightening nut
• Connection box key (provided)
• Input and output cables for wiring
• Small slotted screwdriver for screw terminals
• PTFE tape (wide) for the probe threads
M16x1.5 (1)
M20x1.5 (2)
PTFE
1. Attach cable glands (not provided by Vaisala) to the wiring lead-throughs as required.
Note the cable gland requirements listed in
Guidelines for Safe Use in Hazardous
). For more information on cable gland options and lead-through
Cable Gland Options and Lead-Throughs (page 14)
.
• Use an M16x1.5 gland for the optional external sensor input wiring (Ex ia).
• Use 1 or 2 M20x1.5 glands for wiring the 4 … 20 mA analog output, power supply input
and RS-485 terminals (as required in your application).
• Seal unused lead-throughs.
2. Apply PTFE tape on the probe threads.
• Inspect the threads and remove any possible dirt.
• Wrap 2-3 revolutions of tape in the direction of the thread spiral, starting from the
first thread. Keep the edge of tape parallel to the face of the thread while wrapping.
• Do not tape beyond the edge of the threads or leave loose tape hanging.
The correct thread type in which to install MGP261 is
1.5"
female NPT
. Installing into any other thread type can damage the
equipment and compromise the leak tightness of the connection. If unsure,
verify the thread type with the NPT 1.5" thread test plug.
CAUTION!
MGP261 User Guide
M212273EN-A
26