8.2 Calibration and Adjustment with Insight PC
Software
describes the calibration and adjustment selections in Insight.
1
2
3
4
5
6
7
8
Figure 24 Calibration Menu View in Insight PC Software
1
Calibration information: enter information about the latest calibration (date, text
description) in this tab. Also includes the calibration interval and calibration reminder
selections.
2
Factory calibration: shows read-only information about the factory calibration.
3
Zero point adjustment: adjusts the zero point signal (simultaneously for all gas
measurement parameters). This adjustment replaces the low end adjustment in 2-point
adjustments.
4
Compensation setpoint configuration for the calibration and adjustment environment.
5
Methane (CH
4
) measurement adjustment tab. In 2-point adjustments, use this tab for high
end adjustment and the zero point adjustment for low end adjustment.
6
Carbon dioxide (CO
2
) measurement adjustment tab. In 2-point adjustments, use this tab
for high end adjustment and the zero point adjustment for low end adjustment.
7
Temperature measurement adjustment tab.
8
Adjustment data tabs show the latest stored adjustment for each parameter. Use the scroll
arrows next to the menu tabs on the left and right edge of the view to view all tabs.
1. Connect to Insight (see
Connecting to Insight Software (page 35)
),
MGP261 User Guide
M212273EN-A
52