5. After tightening the probe gland, tighten the 2 screws on the clamp.
A high process pressure can cause the probe to detach from
the adapter if the clamp and gland have been left loose. Always fully
tighten the clamp and probe gland.
WARNING!
6. Attach the mounting plate to the mounting surface with screws or ties. For mounting
7. Check that the probe sits closely in the clamps after mounting. The mounting kit and the
flow-through adapter have been aligned so that the probe sits in the adapter in a straight
angle. A gap between the probe body and clamp indicates that the mounting alignment is
incorrect and must be adjusted.
90°
Ensure the adapter mounting does not pull the probe out of a
straight angle. If the mounting pulls the probe to the side, the leak
tightness of the gland O-ring can be affected.
CAUTION!
8. Connect the gas inlet tube to the upper port of the adapter and the gas outlet tube to the
lower port. Adapter port size: G 1/8.
9. Connect to Vaisala Insight PC software and ensure that the installation type has been set
as
Flow-through adapter
. For instructions, see
Installation Type Configuration in Insight
More Information
‣
‣
Installation Option: Flow-Through Adapter Installation (page 23)
‣
Installation Type Configuration in Insight (page 33)
‣
Flow-Through Adapter Dimensions (page 67)
MGP261 User Guide
M212273EN-A
32