When making a 2-point adjustment, first carry out the zero signal adjustment in
the
Zero point adjustment
tab. The zero signal adjustment replaces the low end
adjustment in a 2-point adjustment.
1. Reset any possible existing adjustments with the
Restore factory adjustment
button. This
prevents earlier adjustments from having an effect on the current adjustment.
2. Insert your device in the reference environment.
3. Before starting the adjustment, make sure that your environmental compensation settings
are correct for your present environment. You can review and change the settings for
pressure and temperature compensation setpoints in the
Compensation setpoints
tab.
4. Wait for the measurement (shown in the graph) to stabilize fully.
5. When the measurement has stabilized, click the
Reference value, point 1
text box and
enter the known CH
4
level of the calibration point. Press
ENTER
or click outside the text
box when done.
6. Check that the measured value for point 1 is automatically inserted.
7. Check the difference between the reference and the measured value. Very large
differences may be due to insufficient stabilization time or unsuitable calibration setup.
a. If you want to apply the adjustment you have made, select
Activate adjustment
.
b. To exit without taking the adjustment in use, select
Close
.
8. After completing the adjustment, update the information in the
Calibration Information
tab.
8.2.4 Calibrating and Adjusting Carbon Dioxide (CO
2
)
Measurement
Prepare the calibration gases required to create the reference condition for the adjustment,
and the setup for controlled calibration gas feed to the probe.
When making a 2-point adjustment, first carry out the zero signal adjustment in
the
Zero point adjustment
tab. The zero signal adjustment replaces the low end
adjustment in a 2-point adjustment.
1. Reset any possible existing adjustments with the
Restore factory adjustment
button. This
prevents earlier adjustments from having an effect on the current adjustment.
2. Insert your device in the reference environment.
3. Before starting the adjustment, make sure that your environmental compensation settings
are correct for your present environment. You can review and change the settings for
pressure and temperature compensation setpoints in the
Compensation setpoints
tab.
4. Wait for the measurement (shown in the graph) to stabilize fully.
Chapter 8 – Calibration
55