6. Analog Output Configuration
6.1 Overview
MGP261 has 3 scalable 4 … 20 mA analog output channels. Each output has the following
configuration options:
• Output parameter selection
• Output scale low end and high end
• Error output level
The configuration of the outputs (output parameter selection and scaling) is selected when
ordering the probe, and can be changed with Modbus or Vaisala Insight PC software.
6.2 Configuring Analog Outputs with Modbus
The Modbus implementation of MGP261 includes configuration registers for analog output
measurement parameter selection, scaling, and error output level.
For a description of MGP261 Modbus registers, see
6.3 Configuring Analog Outputs with Insight
Figure 18 Analog Output Configuration Options in Insight
You can configure the measurement parameter sent on each analog output, the scaling of the
parameter, and the error output level. All 3 outputs have the same configuration options.
MGP261 User Guide
M212273EN-A