A 14-bit image has a maximum grey level of 16383.
-
A 16-bit image has a maximum grey level of 65535
Select the saturation option and the saturated pixels are displayed in red:
Original display
Same image with the saturated pixels indicated
in red
Note: If an image is being acquired and the «Saturation» option is checked, the modification is
applied to the current acquired image
Note: A saturated image creates quantification error when studied by an image analysis software.
Gel-doc systems have to indicate to the user if the image is saturated and if it is then necessary to
modify the integration time.
Select the color palette
The pseudo colors can display different types or levels of fluorescence in an image. It replaces the
original grey levels of the image by another palette color. The FireReader 1D software has several
predefined palette designs. Select your palette design from the followings:
Original image
Coomassie painted image
Original image
GFP painted image