9.GRAPHIC SCREEN
66
RELATIVE Time
The t
ime (ΔT) between
the two cursors shows on the button of the right.
Only the plot on the intersection of the movable cursor with wave displays on the top.
9-2-2. Graphic mode
Select from Single, Continuity or Level.
Single
Only draws the line once to X End Point by pressing
or
external input.
Continuity
Start drawing by pressing
or external input then repeat
drawing continuously until
is pressed or externally inputted.
Level ( ) + exterior
Start the level detection by external input or
.
When the indicated value exceeds the setting value of trigger level, starts the
drawing then draws the line to X End Point once.
Level ( ) + exterior
Start level detection by external input or
.
When the indicated value passes the setting value of trigger level, starts
drawing then draws the line to X End Point once.
Level ( )
When the indicated value exceeds the setting value of trigger level, starts the
drawing then draws the line to X End Point once.
Level ( )
When the indicated value passes the setting value of trigger level, starts the
drawing then draws the line to X End Point once.
Time in this area is displayed.
Cursor
Weight value
at this point
Plot
The graphics drawn is always initialized (cleared) when any setting below is changed.
- Setting in Graph setting
- Weighing code
- Any item of FINAL, OVER and UNDER in the EACH CODE settings
Notice
START
START
STOP
START
START
Содержание F805AT-CK