14.TROUBLE SHOOTING
98
OFL2 (Gross weight > Ca 9 scale division)
Gross weight exceeds Ca 9 scale division.
Decrease the signal coming from the load cell until the normal display returns.
CAL. ERR. 1 (Calibration error)
Zero Calibration must be entered again.
Zero Calibration should always be done before Span Caibration. If CAL. ERR 1 appears after Span
Calibration, Zero Calibration must be entered again.
After Zero Calibration CAL. ERR 1 will disappear.
Scale Capacity is a primary data for the F805AT-CK. You must re-calibrate if
capacity is changed.
DO NOT CHANGE CAPACITY TO RESET OVER SCALE MESSAGE.
Notice
Span adjustment range of F805AT-CK is 0.3 ~ 2.0mV/V.
The display up to capacity at span calibration is guaranteed.
If load cell output is under 0.3mV/V (Calibration error 6) or over 2.0mV/V
(Calibration error 8) when it reached to the capacity, calibration can not been
performed.
Capacity
0
0.3
2.0
Load
Calibration error 6
Calibration error 8
Load cell output
[mV/V]
Span adjustment range