6.FUNCTION SETTING
38
6-1-4. Motion detection
When signal fluctuation is less than the result (range × Min.Scale Div.) and is within the set period,
weight value is stable and
STAB.
display lights.
The two modes of Motion Detection in the F805AT-CK are: Stable Mode and Check Mode.
In stable mode, D1 to D5 in the figure
at the right and the set range are
compared each time A/D conversion is
made, and if at least one of them
exceeds the range, the STABLE signal
turns OFF immediately.
* D1 is the difference between the current
weight value and one-sec-old weight
value.
In checker mode, D1 to D3 in the figure
at the right and the set range are
compared each time A/D conversion is
made, and if at least one of them
exceeds the range, the STABLE signal
turns OFF immediately.
* D1 is the difference between the current
weight value and 0.09-sec-old weight
value.
Setting of motion detect parameter
1) MD MODE
Select the condition from Stable mode or Checker mode in Motion Detection Mode.
2) MD PERIOD
Setting the period for judging the stability of weight (input range: 0.0 ~ 9.9).
3) MD RANGE
Comparing the value (setting value×minimum scale division) with the range of
weight change (input range: 00 ~ 99).
Operation
D1 D2 D3
D4
D5
D1 D2 D3
D4
D5
0.3
0.6
0.8
0.95
1.00sec.
Weight value
Weight value
〈
e.g.1
〉
〈
e.g.2
〉
D1
D2
D3
0.03
0.06
0.09sec.
MODE
OPERATION
→
→
Press
MOTION DITECTION MODE
→
Select STABLE MODE/ CHECKER MODE
OK
→
1)
MODE
OPERATION
→
→
Press
MD (PERIOD)
→
Input 0.0 ~ 9.9
OK
→
2)
MODE
OPERATION
→
→
Press
MD (RANGE)
→
Input 00 ~ 99
OK
→