13.OVER SCALE & ERROR
96
13. OVER SCALE & ERROR
13-1. Over Scale (Weight Error [Pin 46])
* Net weight = Gross weight - Tare
13-2. Calibration Error
13-3. EXC Error
When the voltage of sensing is too low
, “
EXC ERR
”
turns on.
Check the cable between instrument and load cell.
13-4. Priority of Error Display
When plural ERR occured simultaneously, ERR are display following Order.
The input of A/D Converter overflowed
LOAD
Net weight > Net Over set value
OFL1
Gross weight > Ca 9 scale division
OFL2
Gross weight > Gross Over Set value
OFL3
Error item
Alarm Message
Alarm No.
Re-do Zero Calibration.
CAL.ERR.1
1
The initial tare is beyond the zero adjustment range.
CAL.ERR.2
2
The initial tare is minus.
CAL.ERR.3
3
The input Balance weight is beyonde the Capacity.
CAL.ERR.4
4
The Balance weight is
“00000”.
CAL.ERR.5
5
The loadcell output dose not reach the span adjustment
range.
CAL.ERR.6
6
The loadcell output is miunus.
CAL.ERR.7
7
The loadcell output is beyond the span adjustment range.
CAL.ERR.8
8
The weight value is not stable and Calibration stopped.
CAL.ERR.9
9
Priority
ERR. Over Scale
1
2
3
4
5
6
7
CAL.ERR
EXC ERR
-LOAD
LOAD
OFL2
OFL3
OFL1