5.CALIBRATION
36
5-7. Do Secondary Calibration (Equivalent Calibration)
Input the mV/V of load cell corresponds to the input Balance weight to instrument, instrument will
calibrate it following the inputted values.
Operation
1) Press the
button.
2) Select the "CALIBRATION" tag.
3) Press “BALANCE WEIGHT VALUE” tag, key input the weight value then press
.
4) Press “EQUIV. CALIBRATION” tag, key input the corresponded mV/V to the inputted weight
value then press
.
5) Equivalent Calibration is completed while the
changes to
and display the real
weight on the comparison screen.
If the calibration error CAL.ERR.1 is displayed, redo Zero calibration.
5-8. Net Over/ Gross Over
Conditional formula
1) Net Over:
When Net Weight > Net Over set value, OFL 1 is displayed.
2) Gross Over:
When Gross Weight > Gross Over set value, OFL 3 is displayed.
Operation
MODE
OK
OK
CALS
RUN
1)
MODE
CALIBRATION
NET OVER
→
→
2)
MODE
CALIBRATION
PAGE
→
→
Press
Press
Input 0 ~ 99999
→
OK
→
GROSS OVER
→
Input 0 ~ 99999
→
OK
→