SECTION 9
TROUBLESHOOTING
74
9
9.5
Alarm Detection Capabilities
NOTICE
Alarm history data can be read via serial interface, but cannot be read via remote control
connector.
The fault detection functions shown in Table 9-6 "Table of Alarms" and Table 9-7 "Table of
Warnings" are incorporated for protection in the event of a problem with the turbo molecular
pump or power supply unit.
When an error is detected, check the STATUS lamp (Fig. 2-2 (7)) and Refer to Section 9.5.3
"Reset Procedure" for the appropriate remedy.
9.5.1 Movement in Alarm Detection Capabilities (ALARM)
1. STATUS orange lamp (Fig. 2-2 (7)) lights.
2. The remote-control connector "ALARM" signal turn on.
3. The pump start the protective operations shown in Table 9-6 "Table of Alarms".
4. The detection error is recorded in the error log.
9.5.2 Movement in Alarm Detection Capabilities (WARNING)
1. STATUS orange lamp (Fig. 2-2 (7)) flashes.
2. The remote-control connector "WARNING" signal turn on.
3. Pump operation continues.
4. The detection error is recorded in the error log.
9.5.3 Reset Procedure
1. Refer to the Troubleshooting information and eliminate the cause of the problem.
2. Conduct the reset operation.
3. (ALARM) If the problem has been eliminated, the STATUS orange lamp (Fig. 2-2 (7)) goes
out, the "ALARM" remote-control signals (Refer to Table 6-4) turn off after an alarm was
given, the pump rotor decelerates rotational speed.
(WARNING) When the problem is eliminated after a warning occurred, pump operation
continues.
4. If the problem was not completely eliminated, the alarm is not cancelled.
Содержание UTM3400 Series
Страница 2: ...This page is intentionally left blank...
Страница 10: ...viii Location of Label...
Страница 20: ...6 1 1 2 Descriptions This page is intentionally left blank...
Страница 21: ...2 2IDENTIFICATION AND FUNCTION 2 1 Pump Main Unit 2 2 Control Panel 2 3 External I F Panel...
Страница 25: ...3 3CONSTRUCTION AND PRINCIPLE 3 1 Pump Construction 3 2 Principle of Turbo Molecular Pumping 3 3 Controller...
Страница 30: ...16 3 3 3 Controller This page is intentionally left blank...
Страница 36: ...22 4 4 3 Standards Fulfilled This page is intentionally left blank...
Страница 50: ...36 5 5 5 Notes on Transportation This page is intentionally left blank...
Страница 72: ...58 6 6 8 Communication Specifications This page is intentionally left blank...
Страница 73: ...7 7GAS PURGE...
Страница 82: ...68 8 8 6 Turbo Molecular Pump Return Request This page is intentionally left blank...
Страница 95: ...10 10WARRANTY CLAUSES...
Страница 98: ...84 10 This page is intentionally left blank...
Страница 100: ...Index Index 2 This page is intentionally left blank...