9.3
Nothing Happens After an Operation is Made
71
9
9.3
Nothing Happens After an Operation is Made
Table 9-1
Nothing Happens After an Operation is Made
Problem
Possible causes
Corrective action
Section
1
Power ON/OFF switch in
the ON position but the
turbo molecular pump fails
to operate.
Electrical power cable
not properly connected.
Properly connect the electrical power
cable.
5.2
Electrical power outside
power supply unit’s
power range.
Operate within power supply unit’s
power range.
4.1
2
START/STOP button is
pushed but the turbo
molecular pump does not
accelerate.
In REMOTE mode
(POWER lamp lights)
By maintained push, change to
LOCAL mode (POWER lamp blinks)
6.2
Other causes.
Check the STATUS lamp is not lights
in orange. If an alarm is indicated,
correct the malfunction and reset
alarm.
9.5
3
Remote "STOP" signal
active but the turbo
molecular pump does not
accelerate.
In LOCAL mode
(POWER lamp blinks)
By maintained push, change to
REMOTE mode (POWER lamp lights)
6.2
"STOP" signal active.
Deactivate "STOP" signal.
6.7
Other problems.
Check the STATUS orange lamp is not
ON. If an alarm is indicated, correct
the malfunction and reset.
9.5
4
START/STOP button is
pushed but the turbo
molecular pump does not
decelerate.
In REMOTE mode
(POWER lamp lights)
By maintained push, change to
LOCAL mode (POWER lamp blinks)
6.2
5
Remote "STOP" signal
activated but the turbo
molecular pump does not
decelerate.
In LOCAL mode
(POWER lamp blinks)
By maintained push, change to
REMOTE mode (POWER lamp lights)
6.2
Содержание UTM3400 Series
Страница 2: ...This page is intentionally left blank...
Страница 10: ...viii Location of Label...
Страница 20: ...6 1 1 2 Descriptions This page is intentionally left blank...
Страница 21: ...2 2IDENTIFICATION AND FUNCTION 2 1 Pump Main Unit 2 2 Control Panel 2 3 External I F Panel...
Страница 25: ...3 3CONSTRUCTION AND PRINCIPLE 3 1 Pump Construction 3 2 Principle of Turbo Molecular Pumping 3 3 Controller...
Страница 30: ...16 3 3 3 Controller This page is intentionally left blank...
Страница 36: ...22 4 4 3 Standards Fulfilled This page is intentionally left blank...
Страница 50: ...36 5 5 5 Notes on Transportation This page is intentionally left blank...
Страница 72: ...58 6 6 8 Communication Specifications This page is intentionally left blank...
Страница 73: ...7 7GAS PURGE...
Страница 82: ...68 8 8 6 Turbo Molecular Pump Return Request This page is intentionally left blank...
Страница 95: ...10 10WARRANTY CLAUSES...
Страница 98: ...84 10 This page is intentionally left blank...
Страница 100: ...Index Index 2 This page is intentionally left blank...