Superintend IMD
© Copyright Muuntosähkö Oy - Trafox
5
SYSTEM DESCRIPTION
IM-01.MED, PEC-01, TC-01, IC-01, RD-01 and CLT-01 form a system of devices that can
be used to measure and monitor the insulation resistance and capacitance as well as the
continuity of the PE wire of floating electricity networks (Medical IT systems in
accordance with standard IEC 60364-7-710). Of the aforementioned devices, IM-01.MED
is necessary, while the other devices are accessories.
INSTALLATION
PHYSICAL CONNECTION
The devices are connected to the electrical network, which may contain
dangerous voltage. The device may be installed by a trained electrical
engineering professional only. The device contains no user-serviceable parts
and must not be opened. Using the device in violation of these instructions
may compromise safety.
The IM-01.MED unit is the control unit of the system and is installed in the switchboard.
Two IM-01.MED devices may not be installed galvanically in the same network, for
example on the secondary side of the same transformer. The connection is performed as
presented in Figure 1. The installation and wiring should be performed in accordance
with standards IEC 60364 as well as EN 50110. The operating voltage connection of IM-
01.MED must always be equipped with a coupler or a circuit breaker so that the
electricity supply can be disconnected for the duration of maintenance work, for
example. The location of the disconnectors must be clearly marked in the switchboard.
The coupler or line protection switch should also control a relay or contactor, which
separates the measuring wires from the network to be measured. The IM-01.MED device
is equipped with an internal 1 A fuse. In spite of this, the wires of the operating voltage
supply should still be protected with an external fuse. A suitable if the size is, for
example, 6 A. In a DC operating voltage supply, an external Schurter 0001.2503
(T800mA) fuse should be used.
DEVICE MOUNTING
The IM-01.MED, PEC-01 and CLT-01 devices are intended for installation in a DIN TS35
rail in accordance with standard IEC 60715. They are installed by inserting the upper
edge of the DIN TS35 rail in the groove intended for the DIN TS35 rail on the back of the
device and by pushing the bottom edge of the device backward until the retaining latch
clicks into place.
Содержание SUPERINTEND IC-01
Страница 8: ...8 Copyright Muuntosähkö Oy Trafox Superintend IMD http www trafox fi Figure 1 System connection ...
Страница 21: ...Superintend IMD Copyright Muuntosähkö Oy Trafox 21 http www trafox fi ...
Страница 23: ...Superintend IMD Copyright Muuntosähkö Oy Trafox 23 http www trafox fi ...
Страница 28: ...28 Copyright Muuntosähkö Oy Trafox Superintend IMD http www trafox fi ...
Страница 35: ...Superintend IMD Copyright Muuntosähkö Oy Trafox 35 http www trafox fi ...
Страница 37: ...Superintend IMD Copyright Muuntosähkö Oy Trafox 37 http www trafox fi ...
Страница 70: ...70 Copyright Muuntosähkö Oy Trafox Superintend IMD http www trafox fi 36 3 mm 110 mm 62 mm ...