20
© Copyright Muuntosähkö Oy – Trafox
Superintend IMD
The number of the TC-01, IC-01, RD-01 and PEC-01 units configured in the
system and, if needed, the software versions of the units and IM-01.MED.
The manual resistance measurement of the PEC-01 units can be started.
During the manual resistance measurement, test current pulses are fed into
all channels of all PEC-01 units connected to the system. Therefore, the
manual resistance measurement is permitted in medical locations only when
there is no activity in the area to be monitored.
The intensity of the background light can be adjusted to four different levels.
(LOW / MED-LOW / MED-HIGH / HIGH)
Time and date
All fixed alarms can be acknowledged on one go.
The default display is insulation resistance and capacitance. Other parameters and
functions can be viewed by using the and buttons.
The system test is started from the Monitor menu by pressing T and then selecting Y in
the Start test menu and pressing T. This starts the test of the internal measurement
circuit of the IM-01.MED unit. If the test is completed successfully, the screen displays
momentarily the text Test OK; otherwise the text shown in Test FAILED, and an
insulation fault alarm is given to indicate that the insulation resistance can no longer be
measured.
The Monitor menu functions as follows:
Содержание SUPERINTEND IC-01
Страница 8: ...8 Copyright Muuntosähkö Oy Trafox Superintend IMD http www trafox fi Figure 1 System connection ...
Страница 21: ...Superintend IMD Copyright Muuntosähkö Oy Trafox 21 http www trafox fi ...
Страница 23: ...Superintend IMD Copyright Muuntosähkö Oy Trafox 23 http www trafox fi ...
Страница 28: ...28 Copyright Muuntosähkö Oy Trafox Superintend IMD http www trafox fi ...
Страница 35: ...Superintend IMD Copyright Muuntosähkö Oy Trafox 35 http www trafox fi ...
Страница 37: ...Superintend IMD Copyright Muuntosähkö Oy Trafox 37 http www trafox fi ...
Страница 70: ...70 Copyright Muuntosähkö Oy Trafox Superintend IMD http www trafox fi 36 3 mm 110 mm 62 mm ...