7.
When the damper flush is completed, the instrument drains the sheath fluid from
the fluidics lines and internal reservoirs into the waste container.
8.
When prompted, fill a new sample tube with 5 mL of cleaning solution
(10% bleach solution), place it in the sample tube holder, then move the sample
tube holder in the up position.
9.
Empty the sheath fluid and waste containers, then fill the sheath fluid container
with the remaining cleaning solution (~995 mL of 10% bleach solution).
10.
Press
Continue
to begin the first step (decontamination) of the sheath fluidics
cleaning. This step takes about 30 minutes to complete.
11.
The instrument pressurizes the fluidics system, performs one cycle of debubble
and one cycle of fluidic flush, then drains the 10% bleach solution from the
fluidics lines and internal reservoirs into the waste container.