Automated Cell Sorter User Guide
103
Replace peristaltic pump tubing
The peristaltic pump tubing for the sample and sheath pumps is a consumable
part and requires regular replacement for the instrument to continue normal
operation. Additionally, all other tubing in the system can be replaced by a trained
technician, if there are concerns relating to the cleanliness of the tubing or
contamination.
WARNING!
Before replacing any tubing, make sure that the iSort
™
Automated Cell Sorter is powered off and unplugged.
Replacement
schedule for
peristaltic pump
tubing
•
When the sample, sheath, or waste tubing is nearing the end of its life, the
instrument displays a warning message to remind you to replace the tubing.
•
You can also check the remaining tubing life for the sample, sheath, and waste
pumps by selecting the corresponding option in the
Maintenance
Tubing
Life
screen (page 82).