44
iSort
™
Automated Cell Sorter User Guide
7.
When you are finished with your Analysis or Sort, press the
Stop
button.
•
If running Analysis, the sample flow stops and the screen displays the
Analysis Complete dialog.
•
If running Sort, the sample flow and sorting stops, and the screen displays
the Sorting Complete dialog.
Note:
Pressing the
Analysis Complete
or the
Sorting Complete
button also
stops the sample flow and displays the Analysis/Sorting Complete dialog.
When the Analysis/Sort is complete, the instrument saves the data acquired
during the Analysis/Sort in FCS 3.0 format in the specified location.
8.
Press
Rinse
to rinse the fluidic lines of the system (page 45) or press
Skip Rinse
to add another sample for Analysis or Sort.
Note:
We recommend that you rinse the fluidic lines if your next sample is
different from the one that you have just run. The rinse procedure clears the
sample line by back-flushing it with sheath fluid to reduce carryover and
decrease the risk of clogging.
If you decide to skip the rinse, the only other option is to add another sample
and sort again. You will not be able to go back to the Home screen without
rinsing.