54
iSort
™
Automated Cell Sorter User Guide
Fluidic Flush
The Fluidic Flush function washes the fluidic lines of the system with buffer or
sheath fluid to prevent sample carryover between runs.
Required solutions
Sheath fluid (1X PBS, flow cytometry grade)
Run Fluidic Flush
1.
Select
Fluidic Flush
from the
Maintenance
screen.
2.
Place a new tube containing 3 mL of sheath fluid or buffer in the sample tube
holder.
3.
Move the sample tube holder up until it clicks into place, then press
Continue
.
The instrument performs the fluidic flush automatically, which takes
approximately 30 seconds to complete.
4.
When completed, press
Done
to return to the Maintenance screen.