iSort
™
Automated Cell Sorter User Guide
59
2.
When prompted, press
Continue
to begin the sample line cleaning.
3.
When prompted, place a tube containing 3 mL of 10% bleach solution into the
sample tube holder.
4.
Move the sample tube holder up until it stops, then press
Continue
to initiate
the first step of the cleaning process.
5.
When prompted, remove the tube containing the 10% bleach solution from the
sample tube holder, and replace it with a tube containing Attune
™
Wash Solution.
6.
Move the sample tube holder up until it stops, then press
Continue
to initiate
the second step of the cleaning process.