iSort
™
Automated Cell Sorter User Guide
107
8.
Remove the dampener and the sample pump tubing from the instrument.
Install new sample
pump tubing
1.
Attach the new sample pulse dampener into the remaining tubing, then place
the sample pulse damper back into its holder.
2.
Make sure that the sample pulse dampener is upright, then attach the bracket
that secures the sample tube dampener and the fluidic lines to the instrument.