100
iSort
™
Automated Cell Sorter User Guide
6.
While the nozzle tip is in the ultrasonic cleaning bath, loosely screw the small
black cup back on the nozzle assembly, then cover it with the black rubber
protector containing deionized water.
7.
If you have removed the nozzle tip after performing the Shut Down operation
(page 88), store the nozzle tip in the detergent solution at room temperature
until the next time you need to use the instrument.
If you have removed the nozzle tip to clear a clog and wish to resume your
Analysis or Sort, go to Step 8.
8.
Remove the nozzle tip from the detergent solution, then rinse it thoroughly with
deionized water before reinstalling it on the nozzle assembly (page 101).
9.
Optional:
Use a 1-mL syringe to force deionized water backward through the
85-µm nozzle tip to ensure that the nozzle tip is free of obstructions. Repeat as
necessary, switching between the nozzle tip and the larger opposite end to
remove any deposits or clogs.
Note:
Watch the nozzle tip as you force the fluid through it. If clean, the fluid
stream should come out straight from the tip.
10.
When all the deposits or clogs have been removed from the nozzle tip and the
nozzle tip is free of all obstructions, reinstall the nozzle tip (page 101).