iSort
™
Automated Cell Sorter User Guide
77
9.
On the Flow Rate Calibration screen, verify the presence of the performance
tracking beads in the fluorescence scatter plot, which consist of equal
concentrations of beads of four fluorescence emission intensities.
10.
After the instrument automatically adjusts the flow rate calibration parameters,
it displays the “Recalibration Successful” message.
11.
Press
Home
to complete the Flow Rate Calibration procedure.
12.
When prompted, replace the sample tube with the calibration beads with an
empty tube in the tube holder, then press
Continue
to start the final rinse
procedure.
13.
After the rinse is completed, press
Home
to return to the Home screen.