5
•
Installation
Page 78
Vanquish Charged Aerosol Detectors Operating Manual
5.6.5.2
Connecting the Capillary Between Flow Cell and Nebulizer
If the charged aerosol detector is the second detector in the Vanquish
system flow path, for example after the diode array detector, connect
the capillary between flow cell and nebulizer.
Observe the following guidelines for series flow configuration:
•
The charged aerosol detector must be the last detector connected in
the flow path.
•
Other detectors contribute to extra-column volume, which may have
an adverse effect on the peak shape measured by the charged
aerosol detector.
•
If the charged aerosol detector is the second detector in the
Vanquish system after the diode array detector
Be aware of the backpressure limit of the LightPipe™ flow cell in the
diode array detector. Connect the capillary from the flow cell
directly to the nebulizer. Avoid connecting any additional
components in the flow path between the two detectors. Refer to
the
Operating Manual for the diode array detector
.
•
If you want to bypass the flow cell, use the Viper union to
interconnect the capillary from the column with the capillary
connected to the nebulizer. Keep the capillary connection as short as
possible to minimize peak dispersion.
Preparations
•
See
•
Observe all notes and specifications for the flow cell in the first
detector in the Vanquish system. Refer to the
Operating Manual
for
this detector.
Follow these steps
1. Connect the capillary from the column to the flow cell inlet. Refer to
the instructions in the
Operating Manual
for the first detector in the
Vanquish system flow path.
2. Route the connecting capillary from the flow cell outlet upward to
charged aerosol detector.
Содержание Vanquish VF-D20
Страница 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Страница 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...
Страница 182: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...