5
•
Installation
Vanquish Charged Aerosol Detectors Operating Manual
Page 77
Follow these steps
Route and connect the capillary to the nebulizer as required by the
system arrangement:
•
Connecting the capillary between column and nebulizer (see
section 5.6.5.1, page 77
–or–
•
Connecting the capillary between flow cell and nebulizer (see
Follow the respective instructions.
5.6.5.1
Connecting the Capillary Between Column and Nebulizer
If the charged aerosol detector is the only detector in the Vanquish
system flow path, connect the inlet capillary between column and
nebulizer.
Preparations
See
Follow these steps
1. Connect the capillary to the column outlet.
2. Route the inlet capillary from the column compartment through the
guide hole in the detector enclosure to the nebulizer. Use the guide
hole that is next to the column compartment.
3. Connect the inlet capillary to the nebulizer.
Figure 23: Inlet capillary and waste line connected in the detector
Содержание Vanquish VF-D20
Страница 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Страница 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...
Страница 182: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...